Select region
Submit

Best Practices for Multi-Use Bioreactor Sterilization

Topic
Research Laboratories
Woman working on a multi use bioreactor
Topic
Research Laboratories
Subscribe

Stay informed! Subscribe for the latest articles

Best Practices for Multi-Use Bioreactor Sterilization

Regardless of whether the bioreactor is being prepared for a same-use application or an entirely different one, all bioreactor components, as well as liquid media, must be made ready through an effective and reproducible process.

Sterilization Process

Regardless of operation mode, the autoclave must meet the minimum temperature plateau of 121°C with a minimum dwell time of 20 minutes to ensure the sterilization cycle’s efficacy.

3 test types and 3 sterilization methods

  1. Standard Test Cycle

    Steam injection heats to 121°C for a minimum of 20 minutes.

  2. Pre-Vacuum

    Pre-vacuum is added to increase air removal and improve steam penetration within vessel and tubing.

  3. Pre-Vacuum and Pressure Pulses

    Pre-vacuum is followed by steam pressure pulses to remove any remaining air and provide maximum steam penetration in vessel and tubing.

Empty and Full Bioreactor

To evaluate the efficacy of an extended dwell time a test was performed in Getinge Sterilization Test Lab.

Testing compared the results of a 40-minute dwell time in an empty bioreactor (CC 899) with a 3L vessel filled 2.4L (CC 900) for the same dwell time.

evaluation of efficacy of extended dwell time empty and full bioreactor

In CC 899, the temperature sensor placed in the thermowell still did not reach 121°C, even by the end of the 40-minute period. However, bioindicator test results indicated no bacterial growth, verifying that extended plateau time improves sterilization results.

In CC 900, the extended plateau time and reduced liquid in the vessel allowed the minimum temperature to be reached throughout. Bioindicator results also verified sterilization in this scenario.

In CC 900, the extended plateau time and reduced liquid in the vessel allowed the minimum temperature to be reached throughout. Bioindicator results also verified sterilization in this scenario.

Download the application brief here:

Best practices for Multi-use Bioreactors application brief
Get the application brief

Download the application brief

Related Articles

Dear user, You now see International fallback content. Getinge ANZ does not review or control this content. It is possible that some of the products on Getinge international are not approved in your country. For more specific  information relevant for your region please contact us.